Credentials
Laboratory quality policy
Compliance rules
The following are defined as acceptance criteria:
Criterion 1:
If a sample has a content of less than LOD the result is provided as <LOD and the sample is declared compliant with the legislation if the LOD is at least 5 times below the law
Criterion 2:
If the sample has a content within the limits of the legislation, then it is stated that the sample is within the corresponding legislation.
Criterion 3:
If the sample has a content close to the legislation and the uncertainty does not exceed the legislation, then the sample is declared within the legislation.
Criterion 4:
If the sample has a content close to the legislation and the uncertainty exceeds the legislation, then the sample is declared outside the legislation (consumer safety).
Criterion 5:
If the sample has a content greater than the legislation, then it is declared outside the legislation.
Criterion 6:
If it is a customer specification, then if the sample has a content close to the out-of-specification limits and the uncertainty is outside the limits of the customer's specifications, then the customer is informed of the criticality of the material specifications and a decision is made on a case-by-case basis whether to accept the sample.
Criterion 7:
If it is an analysis given to an external provider then the decision rule is in accordance with the legislation or specifications as the laboratory cannot document the uncertainty.
Address
46 Theron Street, Volos
Phone number
2421 305698
Email address
soilabsa@gmail.com
![]()
